Portable capillary electrophoresis system with LED-absorbance photometric and LED-induced fluorescence detection: Design, characterisation and testing
Karlstad University, Division for Chemistry.
2008 (English)Independent thesis Basic level (professional degree), 20 points / 30 hpStudent thesis
Abstract [en]

Capillary electrophoresis (CE) has a wide range of applications in the field of analytical chemistry. In general the most expensive part in a CE system is the detector due to the fact that the detector must have a high sensitivity for small detection volumes and low concentrations. Building portable instruments is one way to make the instruments cheaper and has the advantage that they can be used virtually everywhere. However, downscaling of CE instruments puts some extra demands on the detector. This report describes the design and building of two homemade light-emitting diode (LED) based detectors; a LEDabsorbance photometric detector (LED-AP) and a LED-induced fluorescence (LED-IF) detector. The main goal was to install them inside a portable CE and make a simple separation. The performance of the two detectors had to be evaluated before the main goal could be achieved. p-Nitrophenol was used to create a sensitivity graph for the LED-AP detector, calculating the upper linearity to 5.6 mM when the sensitivity had dropped 10 % caused by non-linearity. The sensitivity graph also showed that the detector had an effective pathlength of 74.2 µm and a stray light of 4.5 % for a 75 µm i.d fused-silica capillary. The LED-IF detector was evaluated by determining the limit of detection (LOD) for fluorescein, at a signal to noise ratio of 3. The LOD was 0.72 µM ± 0.01 µM when immersion oil was used to limit the light scattering from the optic fibres in to the capillary and 0.58 µM ±0.02 µM when silicone oil was used. Without doing any improvements only the LED-AP detector could be used in the portable CE. As a common application area for portable CE instruments is environmental analysis, indirect detection using p-nitrophenol as a probe for separating anions was done to test the system. All analytes were eluted in less than 4 minutes.

Place, publisher, year, edition, pages
2008. , p. 61
Keywords [en]
Capillary electrophoresis, Portable CE, LED-induced fluorescence detection, LED-absorbance photometric detection
National Category
Analytical Chemistry
Identifiers
URN: urn:nbn:se:kau:diva-1364OAI: oai:DiVA.org:kau-1364DiVA, id: diva2:5248
Presentation
2007-09-18
Uppsok
fysik/kemi/matematik
Supervisors
Examiners
Available from: 2008-02-04 Created: 2008-02-04

Open Access in DiVA

fulltext(1527 kB)5245 downloads
File information
File name FULLTEXT01.pdfFile size 1527 kBChecksum SHA-1
c0b9140e10cfbb77b36d5560c517bc3cbef96b83a35217a5dc64274dd7b5d04b0af35cda
Type fulltextMimetype application/pdf

By organisation
Division for Chemistry
Analytical Chemistry

Search outside of DiVA

GoogleGoogle Scholar
Total: 5245 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 802 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf