Laboratory water window cryomicroscopy has recently demonstrated similar image quality as synchrotron-based microscopy but still with much longer exposure times, prohibiting the spread to a wider scientific community. Here we demonstrate high-resolution laboratory water window imaging of cryofrozen cells with 10 s range exposure times. The major improvement is the operation of a lambda = 2.48 nm, 2 kHz liquid nitrogen jet laser plasma source with high spatial and temporal stability at high average brightness >1.5 x 10(12) ph/(s x sr x mu m(2) x line), i.e., close to that of early synchrotrons. Thus, this source enables not only biological x-ray microscopy in the home laboratory but potentially other applications previously only accessible at synchrotron facilities.